ISO 594/1 FIGURE 4 FEMALE REFERENCE CONICAL FITTING
ISO 80369-7:2016cancels and replaces the first editionISO 594-1:1986and the second edition ofISO 594-2:1998, clauses, subclauses, tables, figures, and annexes of which have been consolidated and technically revised.
Compatibility/Interchangeability Between ISO 594/1 Fig 4 and ISO 80369-7 Fig C.5
The ISO 594/1 Figure 4 Female Reference Conical Fitting is replaced byISO 80369-7 Fig C.5Female Reference Luer Slip Connector. The ISO 80369-7 Figure C.5 is not identical to the ISO 594/1 Figure 4 Reference Fitting; so it cannot be certified as ISO 80369-7 Figure C.5. The good news is that the ISO 80369-7 Figure C.5 Female Reference Luer Slip Connector is backward compatible with the ISO 594/1 Figure 4 Reference Fitting; so it can be certified and used in ISO 594/1 designated testing. For those who just need to know the full details; the minimum conical length changed from the ISO 594/1 Figure 4 requirement of 7.5mm minimum; to the ISO 80369-7 Figure C.5 requirement of 10.5mm minimum. I think it is obvious to see that an ISO 80369-7 Figure C.5 Female Reference Luer Slip Connector with a certified minimum length of 10.5 will always meet the ISO 594/1 Figure 4 Female Reference Conical Fitting minimum requirement of 7.5mm minimum.
ISO 594/1 Fig 4 Female Reference Conical Fitting
The ISO 594/1 Fig 4 Female Reference Conical Fitting is used in the ISO 594/1 specified tests of the ISO 594/1 Luer Male 6% Conical Medical Connectors. Keeping with the requirements of ISO 594/1; the ISO 594/1 Fig 4 Female Reference Conical Fitting is made from hardened stainless steel.
The ISO 594/1 Fig 4 Female Reference Conical Fitting has design details which are not specified in ISO 594/1 and which allow connection to test equipment. These design details vary by manufacturer. With the addition of an adapter the various designs will all work with the Enersol Test Equipment; so there is no need to replace existing ISO 594/1 Reference Fittings to utilize the Enersol Test Equipment. The photos above help illuminate these design differences.
The ISO 594/1 Fig 4 Female Reference Conical Fitting is an integral part of Luer needle and syringe testing and is. The ISO 594/1 Fig 4 Female Reference Conical Fitting is assembled to samples for tests such as liquid leakage, air leakage, separation force, and stress cracking.
Uses for the ISO 594/1 Fig 4 Female Reference Conical Fitting
Below is a list of tests which are best accomplished by using the ISO 594/1 Fig 4 Female Reference Conical Fitting.
ISO 594/1 Liquid Leakage Test as defined in paragraph 4.2
To conduct the Liquid Leakage Test found in ISO 594/1 paragraph 4.2 the following Enersol products are required:
S12 - Female Reference Fitting ISO594/1 Figure 4
S15 - Force/Torque Fitting Assembly Machine
S16 - Liquid Leakage Test Machine
ISO 594/1 Air Leakage Test as defined in paragraph 4.3
To conduct the Air Leakage Test found in ISO 594/1 paragraph 4.3 the following Enersol products are required:
S12 - Female Reference Fitting ISO594/1 Figure 4
S15 - Force/Torque Fitting Assembly Machine
S17 - Air Leakage Test Machine
Disclaimer:
This data is provided for general information only. The intention is to provide accurate information; regardless; errors may exist in the supplied information. If accuracy is critical, base your final decisions on the data provided in the root documents: ISO594/1:1986; ISO 594/2:1998 and ISO80369-7:2016; which are copyrighted documents.
To purchase copies visit an Authorized Reseller.
Comments:
Original Posting: 3/20/2009
Last Revision: 8/30/2016
Error corrections in, or comments about, the above data can be sent to: office@gagecrib.com
Gage Crib Worldwide, Inc.
6701 Old 28th St SE, Suite B
Grand Rapids, MI 49546-6937
Phone: 001-616-954-6581 • Fax: 001-616-954-6583 CONTACT FORMS & INFO